[Structural analysis of 16 different coronary stent systems]. / Strukturanalyse von 16 verschiedenen Koronarstentsystemen.
Z Kardiol
; 86(12): 990-9, 1997 Dec.
Article
em De
| MEDLINE
| ID: mdl-9499497
Parallel with the increasing number of coronary stent implantation procedures, new or more elaborate coronary stents were introduced into clinical use almost every month in 1996. We have studied the architecture, surface-morphology, and shape-conversion of 16 different coronary stent systems. The objective of our study was to detect possible specific design and surface features which might influence the healing process in stent restenosis and stent thrombosis. Using electron microscopy we found ultrasmooth surfaces on 5 stents. All other coronary stents revealed individually varying and partly sharp-edged surface irregularities. Video-based morphometry showed covering between 8.3% and 26.4%. Stent shortening after in-vitro expansion ranged from 0% up to 10.5% of the initial stent length. In some stent systems balloon dilation lead to over-expansion of the stent margin. This over-expansion amounted up to 20% of the balloon diameter and the stents revealed a rather biconcave than homogeneous expansion pattern. This phenomenon was also found after postmortal stent implantation in 20 autopsied hearts.
Buscar no Google
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Angioplastia Coronária com Balão
/
Stents
/
Doença das Coronárias
Limite:
Humans
Idioma:
De
Revista:
Z Kardiol
Ano de publicação:
1997
Tipo de documento:
Article
País de publicação:
Alemanha