ABSTRACT
A series of tests were conducted to evaluate the accuracy of a diode detector array in measuring flatness and symmetry of x-ray and electron energies, and the ability to determine change in beam energy. The results show the diode detector array accurately measures flatness and symmetry and can accurately measure or detect change in beam energy. The evaluation also demonstrates the utility of using this model detector system in a linear accelerator QC program.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Aceleradores de Partículas / Radiometria / Eletrônica Médica Tipo de estudo: Diagnostic_studies / Evaluation_studies / Prognostic_studies Limite: Humans Idioma: En Revista: Med Phys Ano de publicação: 1998 Tipo de documento: Article País de afiliação: Estados Unidos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Aceleradores de Partículas / Radiometria / Eletrônica Médica Tipo de estudo: Diagnostic_studies / Evaluation_studies / Prognostic_studies Limite: Humans Idioma: En Revista: Med Phys Ano de publicação: 1998 Tipo de documento: Article País de afiliação: Estados Unidos