ABSTRACT
A simple noninvasive method is described for determining the backscatter to a monitor chamber of a medical accelerator based on the measurement of charge deposited in the target. This method is compared quantitatively to the more elaborate telescopic method for photon beams of 6 MV and 15 MV on linear accelerators having mica and Kapton monitor chambers. The new target charge method gives results consistent with the telescopic method to within 0.3%.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Aceleradores de Partículas / Planejamento da Radioterapia Assistida por Computador / Radioterapia de Alta Energia Limite: Humans Idioma: En Revista: Med Phys Ano de publicação: 1998 Tipo de documento: Article País de afiliação: Estados Unidos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Aceleradores de Partículas / Planejamento da Radioterapia Assistida por Computador / Radioterapia de Alta Energia Limite: Humans Idioma: En Revista: Med Phys Ano de publicação: 1998 Tipo de documento: Article País de afiliação: Estados Unidos