Your browser doesn't support javascript.
loading
Preservation of humidity and heat of respiratory gases in spontaneously breathing, tracheostomized patients.
Thomachot, L; Viviand, X; Arnaud, S; Vialet, R; Albanese, J; Martin, C.
Afiliação
  • Thomachot L; Department of Anesthesia and Intensive Care, Hôpital Nord, Marseilles University Hospital System, Marseilles School of Medicine, France.
Acta Anaesthesiol Scand ; 42(7): 841-4, 1998 Aug.
Article em En | MEDLINE | ID: mdl-9698962
ABSTRACT

BACKGROUND:

Ventilation with endotracheal intubation bypasses the upper airway and the normal heat and moisture exchanging process of inspired gases. A continuous loss of moisture and heat occurs and predisposes patients to serious airway damage. We therefore prospectively studied one heated humidifier system, one cold humidifier system and one heat and moisture exchanger in spontaneously breathing, tracheostomized intensive care unit patients to determine the ability to preserve patients' heat and water.

METHODS:

Following a randomized order, 10 patients were spontaneously ventilated for 24-h periods with the Nam 35 humidifier (Europe Medical, France, a cold water humidifier), the heat and moisture exchange Trach-Vent (Gibeck, Sweden), and the Aerodyne humidifier (Kendall USA). In each patient, during the inspiration phase, the following measurements were performed mean values of temperature and relative humidity of inspired gases. The absolute humidity was calculated. Values were obtained in each patients after 40 min and 24 h.

RESULTS:

The Trach-Vent filter and the Aerodyne humidifier had better humidification and thermic capacities than the Nam 35 humidifier (P < 0.001). With the Nam 35 humidifier, no patient had temperature of inspired gas > 29 degrees C. Concerning absolute humidity of inspired gases, the Nam 35 humidifier achieved a lower performance than the other two tested systems (P < 0.001).

CONCLUSION:

In spontaneously breathing, tracheostomized intensive care unit patients, the Trach-Vent heat and moisture exchanger and the Aerodyne heated system achieved satisfactorily preservation of heat and humidity of inspired gases.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Respiração / Terapia Respiratória / Traqueostomia Tipo de estudo: Clinical_trials / Observational_studies / Prognostic_studies Limite: Female / Humans / Male / Middle aged Idioma: En Revista: Acta Anaesthesiol Scand Ano de publicação: 1998 Tipo de documento: Article País de afiliação: França País de publicação: ENGLAND / ESCOCIA / GB / GREAT BRITAIN / INGLATERRA / REINO UNIDO / SCOTLAND / UK / UNITED KINGDOM
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Respiração / Terapia Respiratória / Traqueostomia Tipo de estudo: Clinical_trials / Observational_studies / Prognostic_studies Limite: Female / Humans / Male / Middle aged Idioma: En Revista: Acta Anaesthesiol Scand Ano de publicação: 1998 Tipo de documento: Article País de afiliação: França País de publicação: ENGLAND / ESCOCIA / GB / GREAT BRITAIN / INGLATERRA / REINO UNIDO / SCOTLAND / UK / UNITED KINGDOM