Your browser doesn't support javascript.
loading
Automated measurement of peak widths for the determination of peak capacity in complex chromatograms.
Lan, K; Jorgenson, J W.
Afiliação
  • Lan K; Department of Chemistry, University of North Carolina at Chapel Hill 27599-3290, USA.
Anal Chem ; 71(3): 709-14, 1999 Feb 01.
Article em En | MEDLINE | ID: mdl-9989387
The peak capacity was measured for an ultrahigh-pressure gradient elution chromatogram of a fluorescently tagged tryptic digest of ovalbumin. The peak widths in the chromatogram were determined by measuring the peak height and the second derivative at the peak maximum. This approach for measuring peak widths was programmed into a computer, and the software accurately determined the general progression of peak widths by measuring 47 peaks throughout the chromatogram in under 10 s. Peak capacity was determined by taking the definite integral of the plot of reciprocal base peak width versus retention time. This calculation of peak capacity is a linear transformation with respect to separation space, so the method is more rigorously accurate than previous methods. The peak capacity for the chromatogram was calculated to be 316.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Processamento de Sinais Assistido por Computador / Cromatografia Idioma: En Revista: Anal Chem Ano de publicação: 1999 Tipo de documento: Article País de afiliação: Estados Unidos País de publicação: Estados Unidos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Processamento de Sinais Assistido por Computador / Cromatografia Idioma: En Revista: Anal Chem Ano de publicação: 1999 Tipo de documento: Article País de afiliação: Estados Unidos País de publicação: Estados Unidos