Plastic bottles as spacers for a pressurized metered-dose inhaler: In vitro characteristics
Kissoon, Niranjan; Teelucksingh, Surujpal; Blake, K. V; Kesser, B; Murphy, S. P; Geller, D.
Afiliação
  • Kissoon, Niranjan; University of Florida Health Sciences Centre, Jacksonville, Fl. USA. e-mail: niranjan.kissoon@jax.ufl.edu
  • Teelucksingh, Surujpal; San Fernando General Hospital, Trinidad and Tobago
  • Blake, K. V; Nemours Children's Clinic, Orlando, Fl. USA
  • Kesser, B; Nemours Children's Clinic, Orlando, Fl. USA
  • Murphy, S. P; Nemours Children's Clinic, Orlando, Fl. USA
  • Geller, D; Nemours Children's Clinic, Orlando, Fl. USA
West Indian med. j ; 50(3): 189-93, Sept. 2001. ilus, tab
Artigo em Inglês | MedCarib | ID: med-304
Biblioteca responsável: JM3.1
Localização: JM3.1; R18.W4
ABSTRACT
Homemade spacer devices are commonly used by children with asthma to improve aerosol deposition from pressurized metered dose inhalers (pMDI); however, the efficacy and efficiency of these devices are not fully characterized. We determined the quality of fine particle fraction (<4.7 um) and ultrafine particle fraction (<3.3 um) of three bottles (from 280 ml to 500 ml) commonly used as spacers in Trinidad and Tobago and compared their performance to the commercially available valved holding chamber (OpT) and pMDI. These data were obtained in vitro using a cascade impactor. All 3 bottles and OpT were similar (p>0.05) in reducing the amount of albuterol emitted as large particles (>4.7 um) to less than 10 ug. The diffreent sized bottles (from 280 ml to 500 ml) produced identical quantities of albuterol in the fine particle and ultrafine particle ranges (p>0.05). All of the sample bottle spacers emitted a higher amount (p<0.002) of fine and ultrafine particles than the OpT and pMDI alone. The OpT resulted in a significantly higher fraction of fine particles (p<0.05) and greater quantity of drug (<0.05) in the ultrafine range as compared to MDI only. The sizes of particles obtained from the bottle spacers are those that have a high probability of reaching the lower airway; however, the clinical relevance of these findings remains to be determined. (AU)
Assuntos
Buscar no Google
Coleções: Bases de dados internacionais Base de dados: MedCarib Assunto principal: Nebulizadores e Vaporizadores / Sistemas de Liberação de Medicamentos Limite: Humanos País/Região como assunto: Caribe Inglês / Trinidad e Tobago Idioma: Inglês Revista: West Indian med. j Ano de publicação: 2001 Tipo de documento: Artigo
Buscar no Google
Coleções: Bases de dados internacionais Base de dados: MedCarib Assunto principal: Nebulizadores e Vaporizadores / Sistemas de Liberação de Medicamentos Limite: Humanos País/Região como assunto: Caribe Inglês / Trinidad e Tobago Idioma: Inglês Revista: West Indian med. j Ano de publicação: 2001 Tipo de documento: Artigo
...