Your browser doesn't support javascript.
loading
An Automatic System for the Delivery of Eye-Drops Using a Microinfusion Pump
Artigo em Coreano | WPRIM (Pacífico Ocidental) | ID: wpr-113169
Biblioteca responsável: WPRO
ABSTRACT

PURPOSE:

To evaluate the efficacy of continuous drug delivery system using microinfusion pump by animal study

METHODS:

Schirmer test was performed after connecting the microinfusion pump at the superior fornix of rabbit eye. The pump was settled to instill artificial tears. We also evaluated the efficacy of continuous 0.1% fluorometholone infusion after alkali burn induced by 1N NaOH.

RESULTS:

Average change in Schirmer tear test value was 12.0mm/5min. The recovery after alkali burn was improved in the eye of continuous infusion of steroid eye drops.

CONCLUSIONS:

Microinfusion pump was thought to be useful as a continuous eye drops delivery system. It would be applicable to infuse artificial tears continuously in severe dry eye patients and to infuse antibiotic eye drops in infectous eye disease patients.
Assuntos

Texto completo: Disponível Base de dados: WPRIM (Pacífico Ocidental) Assunto principal: Soluções Oftálmicas / Lágrimas / Queimaduras / Fluormetolona / Sistemas de Liberação de Medicamentos / Álcalis / Oftalmopatias Limite: Animais / Humanos Idioma: Coreano Revista: Journal of the Korean Ophthalmological Society Ano de publicação: 2003 Tipo de documento: Artigo
Texto completo: Disponível Base de dados: WPRIM (Pacífico Ocidental) Assunto principal: Soluções Oftálmicas / Lágrimas / Queimaduras / Fluormetolona / Sistemas de Liberação de Medicamentos / Álcalis / Oftalmopatias Limite: Animais / Humanos Idioma: Coreano Revista: Journal of the Korean Ophthalmological Society Ano de publicação: 2003 Tipo de documento: Artigo
...