Study on Absorbed Dose Determination of Electron Beam Quality for Cross-calibration with Plane-parallel Ionization Chamber / 의학물리
Article em Ko | WPRIM | ID: wpr-115852
Biblioteca responsável: WPRO
ABSTRACT
Absorbed dose to water based protocols recommended that plane-parallel chambers be calibrated against calibrated cylindrical chambers in a high energy electron beam with R50>7 g/cm2 (E> or =16 MeV). However, such high-energy electron beams are not available at all radiotherapy centers. In this study, we are compared the absorbed dose to water determined according to cross-calibration method in a high energy electron beam of 16 MeV and in electron beam energies of 12 MeV below the cross-calibration quality remark. Absorbed dose were performed for PTW 30013, Wellhofer FC65G Farmer type cylindrical chamber and for PTW 34001, Wellhofer PPC40 Roos type plane-parallel chamber. The cylindrical and the plane-parallel chamber to be calibrated are compared by alternately positioning each at reference depth, zref=0.6R50-0.1 in water phantom. The DW of plane-parallel chamber are derived using across-calibration method at high-energy electron beams of 16, 20 MeV. Then a good agreement is obtained the DW of plane-parallel chamber in 12 MeV. The agreement between 20 MeV and 12 MeV are within 0.2% for IAEA TRS-398.
Assuntos
Palavras-chave
Texto completo: 1 Base de dados: WPRIM Assunto principal: Água / Elétrons Tipo de estudo: Prognostic_studies Idioma: Ko Revista: Korean Journal of Medical Physics Ano de publicação: 2009 Tipo de documento: Article
Texto completo: 1 Base de dados: WPRIM Assunto principal: Água / Elétrons Tipo de estudo: Prognostic_studies Idioma: Ko Revista: Korean Journal of Medical Physics Ano de publicação: 2009 Tipo de documento: Article