Interruption in the supply of breathing gas during general anesthesia due to malposition of the vaporizer: A case report / 대한마취과학회지
Korean Journal of Anesthesiology
; : 270-274, 2010.
Artigo
em Inglês
| WPRIM (Pacífico Ocidental)
| ID: wpr-176336
Biblioteca responsável:
WPRO
ABSTRACT
We report a case of interruption in the supply of breathing gas during general anesthesia caused by malposition of the Drager Vapor 2000(R) vaporizer, which was accidentally tilted and lifted off the Selectatec manifold of the anesthesia machine. Because the patient was an 1-month-old infant, we couldn't check if he had experienced awareness with recall. We emphasize the importance of checking the anesthetic vaporizer after mounting it on the back bar of the anesthesia machine.
Texto completo:
Disponível
Base de dados:
WPRIM (Pacífico Ocidental)
Assunto principal:
Respiração
/
Nebulizadores e Vaporizadores
/
Anestesia
/
Anestesia Geral
Limite:
Humanos
/
Lactente
/
Recém-Nascido
Idioma:
Inglês
Revista:
Korean Journal of Anesthesiology
Ano de publicação:
2010
Tipo de documento:
Artigo