ABSTRACT
We report a case of interruption in the supply of breathing gas during general anesthesia caused by malposition of the Drager Vapor 2000(R) vaporizer, which was accidentally tilted and lifted off the Selectatec manifold of the anesthesia machine. Because the patient was an 1-month-old infant, we couldn't check if he had experienced awareness with recall. We emphasize the importance of checking the anesthetic vaporizer after mounting it on the back bar of the anesthesia machine.
Assuntos

Texto completo: Disponível Base de dados: WPRIM (Pacífico Ocidental) Assunto principal: Respiração / Nebulizadores e Vaporizadores / Anestesia / Anestesia Geral Limite: Humanos / Lactente / Recém-Nascido Idioma: Inglês Revista: Korean Journal of Anesthesiology Ano de publicação: 2010 Tipo de documento: Artigo
Texto completo: Disponível Base de dados: WPRIM (Pacífico Ocidental) Assunto principal: Respiração / Nebulizadores e Vaporizadores / Anestesia / Anestesia Geral Limite: Humanos / Lactente / Recém-Nascido Idioma: Inglês Revista: Korean Journal of Anesthesiology Ano de publicação: 2010 Tipo de documento: Artigo
...