Your browser doesn't support javascript.
loading
Numerical assessment of impeller features of centrifugal blood pump based on fast hemolysis approximation model / 生物医学工程学杂志
Journal of Biomedical Engineering ; (6): 1260-1264, 2014.
Article em Zh | WPRIM | ID: wpr-234419
Biblioteca responsável: WPRO
ABSTRACT
The impeller profile, which is one of the most important factors, determines the creation of shear stress which leads to blood hemolysis in the internal flow of centrifugal blood pump. The investigation of the internal flow field in centrifugal blood pump and the estimation of the hemolysis within different impeller profiles will provide information to improve the performance of centrifugal blood pump. The SST kappa-omega with low Reynolds correction was used in our laboratory to study the internal flow fields for four kinds of impellers of centrifugal blood pump. The flow fields included distributions of pressure field, velocity field and shear stress field. In addition, a fast numerical hemolysis approximation was adopted to calculate the normalized index of hemolysis (NIH). The results indicated that the pressure field distribution in all kinds of blood pump were reasonable, but for the log spiral impeller pump, the vortex and backflow were much lower than those of the other pumps, and the high shear stress zone was just about 0.004%, and the NIH was 0.0089.
Assuntos
Texto completo: 1 Base de dados: WPRIM Assunto principal: Circulação Assistida / Estresse Mecânico / Coração Auxiliar / Hemólise / Modelos Teóricos Limite: Humans Idioma: Zh Revista: Journal of Biomedical Engineering Ano de publicação: 2014 Tipo de documento: Article
Texto completo: 1 Base de dados: WPRIM Assunto principal: Circulação Assistida / Estresse Mecânico / Coração Auxiliar / Hemólise / Modelos Teóricos Limite: Humans Idioma: Zh Revista: Journal of Biomedical Engineering Ano de publicação: 2014 Tipo de documento: Article