Lateral flow immunoassay for quantitative detection of ractopamine in swine urine / 生物医学与环境科学(英文)
Biomed. environ. sci
; Biomed. environ. sci;(12): 134-137, 2014.
Article
em En
| WPRIM
| ID: wpr-247072
Biblioteca responsável:
WPRO
ABSTRACT
A strip reader based lateral flow immunoassay (LFIA) was established for the rapid and quantitative detection of ractopamine (RAC) in swine urine. The ratio of the optical densities (ODs) of the test line (AT) to that of the control line (AC) was used to effectively minimize interference among strips and sample variations. The linear range for the quantitative detection of RAC was 0.2 ng/mL to 3.5 ng/mL with a median inhibitory concentration (IC50) of 0.59 ± 0.06 ng/mL. The limit of detection (LOD) of the LFIA was 0.13 ng/mL. The intra-assay recovery rates were 92.97%, 97.25%, and 107.41%, whereas the inter-assay rates were 80.07%, 108.17%, and 93.7%, respectively.
Texto completo:
1
Base de dados:
WPRIM
Assunto principal:
Fenetilaminas
/
Fitas Reagentes
/
Suínos
/
Urina
/
Imunoensaio
Tipo de estudo:
Diagnostic_studies
Limite:
Animals
Idioma:
En
Revista:
Biomed. environ. sci
Ano de publicação:
2014
Tipo de documento:
Article