Fracture of Self-Expandable Metal Stent during Endoscopic Removal in Benign Biliary Stricture
Clinical Endoscopy ; : 95-97, 2013.
Article em En | WPRIM | ID: wpr-28642
Biblioteca responsável: WPRO
ABSTRACT
The endoscopic insertion of the self-expandable metal stent (SEMS) in benign biliary stricture has become an alternative to surgery. Fracture or migration of SEMS can occur rarely as complications. We report a case of fracture of SEMS during endoscopic retrieval in patients with chronic pancreatitis. In this case, broken stent was successfully removed with endoscopic ballooning of bile duct and with a snare device.
Assuntos
Palavras-chave
Texto completo: 1 Base de dados: WPRIM Assunto principal: Ductos Biliares / Stents / Constrição Patológica / Pancreatite Crônica / Proteínas SNARE Limite: Humans Idioma: En Revista: Clinical Endoscopy Ano de publicação: 2013 Tipo de documento: Article
Texto completo: 1 Base de dados: WPRIM Assunto principal: Ductos Biliares / Stents / Constrição Patológica / Pancreatite Crônica / Proteínas SNARE Limite: Humans Idioma: En Revista: Clinical Endoscopy Ano de publicação: 2013 Tipo de documento: Article