Defibrillation efficacy of a newly developed automated external defibrillator in a swine ventricular tachycardia/fibrillation model / 中华心血管病杂志
Zhonghua xinxueguanbing zazhi
; (12): 845-848, 2009.
Article
em Zh
| WPRIM
| ID: wpr-323965
Biblioteca responsável:
WPRO
ABSTRACT
<p><b>OBJECTIVE</b>The aim of this study was to evaluate the recognition and defibrillation efficiency of a newly developed automated external defibrillator (AED).</p><p><b>METHOD</b>Ventricular tachycardia (VT)/ventricular fibrillation (VF) was induced by alternating current (50 Hz) through an electrode placed on apex of right ventricle in 23 anesthetized swine and recorded, recognized and defibrillated by a newly developed AED.</p><p><b>RESULTS</b>A total of 96 VF was induced and 145 defibrillations were recorded. We analyzed available 167 electrocardiosignal with a total length of 103,740 seconds. The accuracy, sensitivity and the specificity of the AED on VT/VF recognition are 99.5%, 98.2% and 99.6%, respectively. The success rate of defibrillation was 33.4% which increased in proportion to defibrillation energy. The defibrillation threshold of energy is 29.10-116.91 (78.75 +/- 35.64) J, the defibrillation threshold of electric quantity is (0.11 +/- 0.04) C and the defibrillation threshold of voltage is (1216.67 +/- 260.87) V.</p><p><b>CONCLUSIONS</b>This newly developed AED has high sensitivity and the specificity on recognizing VT/VF. The lower success rate of defibrillation of this AED is associated with the low defibrillation energy during defibrillation which needs to be improved on further development.</p>
Texto completo:
1
Base de dados:
WPRIM
Assunto principal:
Suínos
/
Terapêutica
/
Fibrilação Ventricular
/
Cardioversão Elétrica
/
Sensibilidade e Especificidade
/
Desfibriladores
/
Modelos Animais de Doenças
/
Desenho de Equipamento
Tipo de estudo:
Diagnostic_studies
Limite:
Animals
Idioma:
Zh
Revista:
Zhonghua xinxueguanbing zazhi
Ano de publicação:
2009
Tipo de documento:
Article