Your browser doesn't support javascript.
loading
A Simple and Rapid Method Based on Liquid Chromatography-Tandem Mass Spectrometry for the Measurement of alpha-L-Iduronidase Activity in Dried Blood Spots: An Application to Mucopolysaccharidosis I (Hurler) Screening
Artigo em Inglês | WPRIM (Pacífico Ocidental) | ID: wpr-34577
Biblioteca responsável: WPRO
ABSTRACT

BACKGROUND:

We developed an analytical method to measure alpha-L-iduronidase (IDUA) activity in dried blood spots. This was achieved by using liquid chromatography coupled to tandem mass spectrometry (LC-MS/MS) with electrospray ionization in the positive ion mode.

METHODS:

Chromatographic separation was completed using mobile phase involving water-formic acid and acetonitrile-formic acid over 2.8 min of run time on a column with a Kinetex XB-C18 (Phenomenex, USA). The detection of column effluent was performed using a Xevo TQ-S triple quadrupole mass spectrometer (Waters, USA) in the multiple-reaction monitoring mode. This method was verified with blank and control samples at four activity levels base, low, medium, and high. Control materials were provided from Centers for Disease Control and Prevention (CDC).

RESULTS:

Intra- and inter-day precisions were between 2.6% and 16.5% and between 7.9% and 17.0%, respectively. A correlative regression study on the IDUA activity in CDC-control samples performed to assess the validity of the developed method showed a highly significant linear association (r2=0.9976) between the calculated and CDC-reported values and an obvious difference in activity among the four levels. This reliable analytical method was applied to mucopolysaccharidosis I (Hurler) screening of patients under treatment (n=4) and in normal controls (n=129). IDUA activity ranged from 8.98 to 77.12 micromol/hr/L) in normal controls, and patients undergoing medical treatment showed low IDUA activity.

CONCLUSIONS:

This method had advantages of simplicity, rapid sample preparation, and liquid chromatographic separation, which efficiently inhibited ionization suppression induced by matrix effects in mass spectrometric detection.
Assuntos

Texto completo: Disponível Base de dados: WPRIM (Pacífico Ocidental) Assunto principal: Especificidade por Substrato / Análise de Regressão / Cromatografia Líquida de Alta Pressão / Mucopolissacaridose I / Espectrometria de Massas em Tandem / Teste em Amostras de Sangue Seco / Iduronidase Tipo de estudo: Estudo diagnóstico / Estudo de rastreamento Limite: Humanos Idioma: Inglês Revista: Annals of Laboratory Medicine Ano de publicação: 2015 Tipo de documento: Artigo
Texto completo: Disponível Base de dados: WPRIM (Pacífico Ocidental) Assunto principal: Especificidade por Substrato / Análise de Regressão / Cromatografia Líquida de Alta Pressão / Mucopolissacaridose I / Espectrometria de Massas em Tandem / Teste em Amostras de Sangue Seco / Iduronidase Tipo de estudo: Estudo diagnóstico / Estudo de rastreamento Limite: Humanos Idioma: Inglês Revista: Annals of Laboratory Medicine Ano de publicação: 2015 Tipo de documento: Artigo
...