Application of ECToolbox software in evaluation of left ventricular ejection fraction / 中国医学影像技术
Article em Zh | WPRIM | ID: wpr-473365
Biblioteca responsável: WPRO
ABSTRACT
Objective To evaluate the applicability of formula R0, R1, R2 of ECToolbox software in measurement of left ventricular ejection fraction (LVEF) with gated myocardial imaging. Methods Rest gated myocardial SPECT with ~(99m)Tc-MIBI and rest gated blood-pool SPECT with ~(99m)Tc-RBC were performed on 64 patients (44 with CHD, 20 with hypertension). LVEFs were separately calculated with formula R0, R1, R2 of ECToolbox software, and the results were recorded as R0 LVEF, R1 LVEF and R2 LVEF, respectively, and compared with LVEFs from gated blood-pool imaging (GBPI LVEF). Results The LVEFs from formula R0, R1 and R2 were all significantly correlated with GBPI LVEF (r=0.905, 0.905 and 0.903, P0.05). Conclusion The formula R1 of ECToolbox software is most applicable in measuring LVEF with gated myocardial imaging.
Palavras-chave
Texto completo: 1 Base de dados: WPRIM Idioma: Zh Revista: Chinese Journal of Medical Imaging Technology Ano de publicação: 2010 Tipo de documento: Article