ABSTRACT
We report the case of a polyethylene tibial post fracture in a 72-year-old woman 14 months after a Scorpio posterior-stabilized (PS) total knee arthroplasty. The polyethylene wear was found around the fracture site of the post, especially over the anterior aspect of the post base. The failure mechanism of the post fracture in the present case was anterior impingement with excessive wear over the base of the anterior aspect of the tibial post, which became a stress-riser of post and cam articulation. This is the first report of a polyethylene tibial post fracture of a Scorpio PS prosthesis.
Assuntos
Palavras-chave
Texto completo: 1 Base de dados: WPRIM Assunto principal: Polietilenos / Reoperação / Falha de Prótese / Artroplastia do Joelho / Osteoartrite do Joelho / Articulação do Joelho / Prótese do Joelho Limite: Aged / Female / Humans Idioma: En Revista: Clinics in Orthopedic Surgery Ano de publicação: 2009 Tipo de documento: Article
Texto completo: 1 Base de dados: WPRIM Assunto principal: Polietilenos / Reoperação / Falha de Prótese / Artroplastia do Joelho / Osteoartrite do Joelho / Articulação do Joelho / Prótese do Joelho Limite: Aged / Female / Humans Idioma: En Revista: Clinics in Orthopedic Surgery Ano de publicação: 2009 Tipo de documento: Article