Dosimetric effect of patient arm position on Cyberknife radiosurgery for spinal tumors / 北京大学学报(医学版)
Artigo em Chinês | WPRIM (Pacífico Ocidental) | ID: wpr-936132
Biblioteca responsável: WPRO
ABSTRACT
OBJECTIVE@#To assess the potential dosimetric effects of arms movement in patients with Cyberknife spine tumors.@*METHODS@#In the study, 12 patients with thoracic and lumbar tumors were retrospectively selected respectively. The contour of the patient's arms was sketched and the CT density was modified to be equivalent to air in order to simulate the extreme case when the arm was completely removed from the radiation fields. The dose of simulated plan was re-calculated with the original beam parameters and compared with the original plan. The changes of V100, D95, and D90, conformity index (CI) and heterogeneity index (HI) in planning target volume (PTV), as well as Dmax, D1cc and D2cc in the spinal cord, stomach, esophagus, and intestines were analyzed by comparing with the original plans.@*RESULTS@#Compared with the original treatment plan, V100, D95, D90 and CI of PTV for the simulated plan was increased by 0.86%, 2.02%, 1.97% and 0.80% respectively, the difference was statistically significant (P < 0.05). Dmax, D1cc and D2cc of spinal cord was increased by 2.35%, 0.59% and 1.49% on average, compared with the original plan, the difference was statistically significant (P < 0.05). The difference was statistically significant only in average D2cc of stomach, which was increased by 1.70%, compared with the original plan (P < 0.05). There was no significant difference in dose change of eso-phagus and intestine between the original and simulated plans.@*CONCLUSION@#This study analyzed the most extreme arm position in spinal tumor of radiation therapy based on Cyberknife. It was found that the change of arm position had little effect on dosimetry. In addition, with the change of arm position, the dose in PTV and organ at risk (OAR) increased, but the increase was relatively small. Therefore, in some special cases where the patient really can't keep the arm position consistent during treatment, reasonable adjustment can be accepted. However, in order to ensure accurate radiotherapy, patient position should be as stable and consistent as possible.
Assuntos

Texto completo: Disponível Base de dados: WPRIM (Pacífico Ocidental) Assunto principal: Braço / Dosagem Radioterapêutica / Neoplasias da Coluna Vertebral / Planejamento da Radioterapia Assistida por Computador / Estudos Retrospectivos / Radiocirurgia / Radioterapia de Intensidade Modulada Tipo de estudo: Estudo observacional Limite: Humanos Idioma: Chinês Revista: Journal of Peking University(Health Sciences) Ano de publicação: 2022 Tipo de documento: Artigo
Texto completo: Disponível Base de dados: WPRIM (Pacífico Ocidental) Assunto principal: Braço / Dosagem Radioterapêutica / Neoplasias da Coluna Vertebral / Planejamento da Radioterapia Assistida por Computador / Estudos Retrospectivos / Radiocirurgia / Radioterapia de Intensidade Modulada Tipo de estudo: Estudo observacional Limite: Humanos Idioma: Chinês Revista: Journal of Peking University(Health Sciences) Ano de publicação: 2022 Tipo de documento: Artigo
...