Clinical Application and Mechanical Analysis of the Treatment of Multiple Rib Fractures and Flail Chest Using Three-wings Rib Plate / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation
; (6): 395-398, 2020.
Article
em Zh
| WPRIM
| ID: wpr-942748
Biblioteca responsável:
WPRO
ABSTRACT
This research evaluated the clinical efficacy of three-wings rib plate in the treatment of multiple rib fractures and flail chest with mechanical analysis and clinical verification. The model of rib and three-wings rib plate was reconstructed. The contact simulation with pretension stress was applied to the plate's fixation, and it was found that the bearable stress of the rib fractures after fixation increased from the result which indicated a good fixation efficacy of the plate. Clinical data of 53 cases of rib fractures and flail chest treated with three-wings rib plate in Shanghai Pudong Hospital of Fudan University were retrospectively analyzed. After the operation, the pain of the patients was relieved. Postoperative CT reconstruction of the chest showed good restoration of the rib fractures, which verified the clinical efficacy of three-wings rib plate. The three-wings rib plate showed a high value in clinical use for treatment of rib fractures.
Palavras-chave
Texto completo:
1
Base de dados:
WPRIM
Assunto principal:
Fraturas das Costelas
/
Costelas
/
China
/
Estudos Retrospectivos
/
Tórax Fundido
/
Fixação Interna de Fraturas
Tipo de estudo:
Observational_studies
Limite:
Humans
País/Região como assunto:
Asia
Idioma:
Zh
Revista:
Chinese Journal of Medical Instrumentation
Ano de publicação:
2020
Tipo de documento:
Article