1.
Rev Sci Instrum ; 89(4): 043706, 2018 Apr.
Artigo em Inglês | MEDLINE | ID: mdl-29716370

RESUMO

For many scientific questions gaining three-dimensional insight into a specimen can provide valuable information. We here present an instrument called "tOMography Nano crYo (OMNY)," dedicated to high resolution 3D scanning x-ray microscopy at cryogenic conditions via hard X-ray ptychography. Ptychography is a lens-less imaging method requiring accurate sample positioning. In OMNY, this in achieved via dedicated laser interferometry and closed-loop position control reaching sub-10 nm positioning accuracy. Cryogenic sample conditions are maintained via conductive cooling. 90 K can be reached when using liquid nitrogen as coolant, and 10 K is possible with liquid helium. A cryogenic sample-change mechanism permits measurements of cryogenically fixed specimens. We compare images obtained with OMNY with older measurements performed using a nitrogen gas cryo-jet of stained, epoxy-embedded retina tissue and of frozen-hydrated Chlamydomonas cells.


Assuntos
Imageamento Tridimensional/instrumentação , Microscopia/instrumentação , Tomografia por Raios X/instrumentação , Animais , Chlamydomonas , Criopreservação/instrumentação , Desenho de Equipamento , Interferometria/instrumentação , Lasers , Imagem Óptica/instrumentação , Retina/citologia , Espalhamento a Baixo Ângulo , Temperatura , Difração de Raios X/instrumentação