Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros

Base de dados
Ano de publicação
Tipo de documento
Intervalo de ano de publicação
1.
Biotechnol Prog ; 37(5): e3187, 2021 09.
Artigo em Inglês | MEDLINE | ID: mdl-34164947

RESUMO

Protein concentration determination is a necessary in-process control for the downstream operations within biomanufacturing. As production transitions from batch mode to an integrated continuous bioprocess paradigm, there is a growing need to move protein concentration quantitation from off-line to in-line analysis. One solution to fulfill this process analytical technology need is an in-line index of refraction (IoR) sensor to measure protein concentration in real time. Here the performance of an IoR sensor is evaluated through a series of experiments to assess linear response, buffer matrix effects, dynamic range, sensor-to-sensor variability, and the limits of detection and quantitation. The performance of the sensor was also tested in two bioprocessing scenarios, ultrafiltration and capture chromatography. The implementation of this in-line IoR sensor for real-time protein concentration analysis and monitoring has the potential to improve continuous bioprocess manufacturing.


Assuntos
Anticorpos Monoclonais/análise , Reatores Biológicos , Proteínas Recombinantes/análise , Refratometria/métodos , Animais , Células CHO , Cromatografia , Cricetinae , Cricetulus , Humanos , Ultrafiltração
SELEÇÃO DE REFERÊNCIAS
Detalhe da pesquisa