ABSTRACT

When analyzing samples containing paraproteins, various interference effects are encountered in the clinical laboratory. Precipitation of paraproteins mostly interferes with the assays that use photometric detection. Herein, we present a case of a patient with multiple myeloma who had paraproteins and spuriously elevated total bilirubin levels (31.1 mg/dL), which were measured by using Roche total bilirubin assay on the Modular DPE (Roche Diagnostics, Switzerland) chemical analyzer. The total bilirubin concentration reduced from 31.1 mg/dL to 1.5 mg/dL, when tested after three fold dilution of the sample on Modular DPE chemical analyzer.


Subject(s)
Humans , Bilirubin , Hyperbilirubinemia , Multiple Myeloma , Paraproteinemias , Paraproteins